ZXH-550 Technical Characteristics

The ventilator is aerodynamic, electronic-controlled ventilator, it has over ten respiratory mode, functions can be set through the respiratory rate, inspiratory time, tidal volume parameter to achieve management of the patient breathe, parameters, Germanic Baode  proportion electromagnetism valve is suction control valve, Japanese SMC electronic air and oxygen mixer and etc. These high quality imported parts bring out the machine is more accurate, more reliable and more stable.

Main technical parameters and performance

1.Respiratory Mode: A/C, SIMV, PSV, SPONT, MMV, IPPV, SIGH, ,PEEP

2.Parameter setting up:

2.1Tidal volume (Vt): 20 ~ 2000ml

2.2Breathing frequency (f): 4 to 99 times / minute

2.3SIMV Frequency: 6 ~ 60 times / minute

2.4Inspiratory time (Ti): 0.2 ~ 6 s

2.5Breath-hold time (Tp): 0 to 2 s

2.6Trigger pressure (Pt): -2kPa~ 2kPa,

2.7PEEP pressure: 0.1 ~ 2kPa,

2.8PSV pressure: 0~ 5 kPa

2.9Safety largest gas pressure: ≤6 kPa

2.10Breathing oxygen concentration: 40 ~ 100%, consecutive adjustable

3 Alarm Performances

3.1Pressure upper limit (Ph) alarm: 2 ~ 6kPa, sound and light alarm.

3.2Pressure lower limit (Pl) alarm: 0 ~ 2kPa, sound and light alarm

3.3Ventilation volume upper limit (MVh) alarm: more than settings, sound and light alarm.

3.4Ventilation volume lower limit (MVl) alarm: less than setting, sound and light alarm.

3.5Power disconnection warning: immediate alarm if power disconnection, warning time≥120s.

3.6quiet sound alarm: ≤2min

4.Overall performance: 

4.1Drive mode: aerodynamic, electronic-controlled

4.2Power supply: AC 110 V --220V. 50/60Hz

4.3Aerodynamic source pressure: 0.28~0.6MPa

4.4mainframe input power: 65VA.

4.5complete machine noise: ≤65 dB (A).

4.6Compliance system: ≤4 ml/100Pa

ZXH-600-C

The ventilator is aerodynamic, electronic-controlled ventilator, it has over ten respiratory mode, functions can be set through the respiratory rate, inspiratory time, tidal volume parameter to achieve management of the patient breathe,  There are Japanese Sharp large screen true color LCD to show waveform  and parameters, Germanic Baode  proportion electromagnetism valve is suction control valve, Japanese SMC electronic air and oxygen mixer and etc. These high quality imported parts bring out the machine is more accurate, more reliable and more stable.

Main technical parameters and performance

1.Respiratory Mode: A/C, SIMV, PSV, SPONT, MMV, IPPV, SIGH, ,PEEP

2.Parameter setting up:

2.1Tidal volume (Vt): 20 ~ 2000ml

2.2Breathing frequency (f): 4 to 99 times / minute

2.3SIMV Frequency: 6 ~ 60 times / minute

2.4Inspiratory time (Ti): 0.2 ~ 6 s

2.5Breath-hold time (Tp): 0 to 2 s

2.6Trigger pressure (Pt): -2kPa~ 2kPa,

2.7PEEP pressure: 0.1 ~ 2kPa,

2.8PSV pressure: 0~ 5 kPa

2.9Safety largest gas pressure: ≤6 kPa

2.10Breathing oxygen concentration: 40 ~ 100%, consecutive adjustable

3Alarm Performances

3.1Pressure upper limit (Ph) alarm: 2 ~ 6kPa, sound and light alarm.

3.2Pressure lower limit (Pl) alarm: 0 ~ 2kPa, sound and light alarm

3.3Ventilation volume upper limit (MVh) alarm: more than settings, sound and light alarm.

3.4Ventilation volume lower limit (MVl) alarm: less than setting, sound and light alarm.

3.5Power disconnection warning: immediate alarm if power disconnection, warning time≥120s.

3.6quiet sound alarm: ≤2min